miniGC Method Note – Residual Solvents

miniGC Method Note – Residual Solvents

Analysis of Residual Solvents standard in the miniGC

miniGC Method

Column MXT-624, 30 m, 0.25 mm ID, 1.40 µm (cat.# 70968-LUC25)
Sample  
   
Injection  
Inj. Vol.: 1.0 µL split (split ratio 10:1)
Liner: 4 mm low pressure drop precision inlet liner w/ wool (cat.# 23309)
Inj. Temp.: 280 °C
Oven  
Oven Temp.: Hold at 35 °C for 7.0 min, Ramp to 215 °C at 20 °C/min, Hold for 2.0 min
Carrier Gas He, constant pressure
Linear Velocity: 40.4 cm/sec (2.50mL/min)
Detector FID @ 320 °C
Instrument Lucidity miniGC

Results

Figure 1: Injection 1
Figure 2: Injection 1
Figure 3: Injection 1

 

More results from the miniGC can be found here: https://luciditysystems.com/products/minigc/minigc-results/

More information on the miniGC can be found here: https://luciditysystems.com/products/minigc/

To see the miniGC in operation go here: https://luciditysystems.com/products/minigc/minigc-interface/