miniGC App Note – Residual Solvents

miniGC App Note – Residual Solvents

Analysis of Residual Solvents standard in the miniGC

miniGC Method

ColumnMXT-624, 30 m, 0.25 mm ID, 1.40 µm (cat.# 70968-LUC25)
Sample
Injection
Inj. Vol.:1.0 µL split (split ratio 10:1)
Liner:4 mm low pressure drop precision inlet liner w/ wool (cat.# 23309)
Inj. Temp.:280 °C
Oven
Oven Temp.:Hold at 35 °C for 7.0 min, Ramp to 215 °C at 20 °C/min, Hold for 2.0 min
Carrier GasHe, constant pressure
Linear Velocity:40.4 cm/sec (2.50mL/min)
DetectorFID @ 320 °C
InstrumentLucidity miniGC

Results

Figure 1: Injection 1
Figure 2: Injection 1
Figure 3: Injection 1

More results from the miniGC can be found here: https://luciditysystems.com/products/minigc/minigc-results/

More information on the miniGC can be found here: https://luciditysystems.com/products/minigc/

To see the miniGC in operation go here: https://luciditysystems.com/products/minigc/minigc-interface/