Lucidity LC-UV Helps With Sleep

Lucidity LC-UV Helps With Sleep

For some time I have been having trouble falling asleep and staying asleep. To combat this I have been trying different things, from exercise to sleep aides.  One of the sleep aides that I have been using is over the counter Melatonin in both pill and gummy form. I wondered about this melatonin and if the label claims on the concentration were correct.

I decided that the improved Lucidity LC-UV would be the perfect instrument for this analysis. The variable injection split loop design will allow for me to make a single standard and use the variable injection to create my calibration curve.  The method used for analysis is below:

Lucidity LC-UV Method
  Flow Rate 1.0 mL/min
  Column Raptor C18 150 x 4.6 mm, 5.0 μm
  Oven Temperature 30 ºC
  Sample injection  10 μL
  Wavelength 240 ±2 nm
  Mobile Phase A Methanol
  Mobile Phase B Water
Gradient
  Time (min) %B
  0.0 70
 10 70

For the experiment I made a standard by dissolving 0.005 g of powdered melatonin in 50 mL of methanol. Using this standard I injected 3 different amounts 5 μL, 10 μL, and 15 μL.  These three injections gave me a concentration of 0.5 mg/mL, 1.0 mg/mL and 1.5 mg/mL of melatonin. 

Overlay of the three standard injections

Overlay of the three melatonin standard injections

Calibration curve of melatonin

Calibration curve of the Melatonin standard

Once the standards were run and the calibration curve made, it was time to run the samples.  I chose two samples to test. They were chosen as the most popular forms of melatonin, one pill and one gummy.

For the samples I decided to dissolve them the same way that I made the standards, with methanol. Neither sample went into solution with just methanol, so I decided to use a mixture similar to the mobile phase used in the method. The solution chosen was a 1:1 methanol water, just a bit different from the mobile phase, but worked well. 

The pill went into solution with 10 minutes of sonication at 30 ºC.

The pill label showed a concentration of 5 mg per pill and to keep it in line with the calibration curve, I added two pill to the 100 mL flask, giving a  concentration of 0.1 mg/mL and injected 10 μL of the sample into the LC-UV.  This should give an overall peak concentration of 1.0 mg/mL.

Chromatogram of melatonin from a pill

Melatonin chromatogram from the pill sample.

As shown in the chromatogram above, the concentration was calculated out to be 1.024 mg/mL. Taking into account the 100 mL dilution, the two pills and the injection volume of 10 μL, the concentration is 5.12 mg per pill, a 102.4% recovery of the label claim.

I attempted to dissolve the gummy with the same solution as the pill. So, one gummy was added to a 100 mL flask and again, the sonicator was set to 30ºC for 10 minutes, but after that time the gummy was still intact. I decided to set the time for 1 hour and see what happened then, after that hour the gummy had changed from red to a white color but was still very much intact. I then set the sonicator for 5 more hours and let it go, and after over 6 hours of sonication, the gummy remained unchanged.

Melatonin gummy in a flask

Melatonin gummy after 6 hours in a sonicator

I then decided to use the Lucidity SimplePrep. The gummy sample was added to a 50 mL centrifuge tube and a flat Teflon coated magnet is then added to the tube. The tube is covered with a Teflon cover and placed into a heating block in the SimplePrep. I programmed the instrument to add 35 mL of the 1:1 methanol:water solution, heat to 33 ºC, and agitate for 60 minutes. Once the instrument was programmed I pressed start and walked away, an hour later the sample was in solution.

Simpleprep with one sample

Lucidity SimplePrep with melatonin gummy sample

Once the hour was up, I checked the sample and the sample was completely in solution, showing that the SimplePrep is a great tool for putting gummy samples into solution.

Two gummy samples side by side

The two samples side by side

I transferred the solution to a 100 mL flask and diluted to the mark, pulled an aliquot for injection on the Lucidity LC-UV. The label for the gummy showed a concentration of 20 mg and a serving size of 2 gummies, therefore the concentration of one gummy should be around 10 mg. With that in mind, adding that to 100 mL flask and an injection volume of 10 μL, I should get a concentration close to 1 mg/mL once again.

Chromatogram of the gummy sample

Chromatogram of the gummy sample

The chromatogram above shows the concentration was calculated out to be 1.083 mg/mL. Taking into account the 100 mL dilution, and the injection volume of 10 μL, the concentration is 10.83 mg per pill, a 108.3% recovery of the label claim.

Overall, the Lucidity LC-UV shows to be very capable of running and testing for the amount of melatonin in various samples. The improved, variable injection volume of the split loop make creating a calibration curve from a single sample easy and intuitive.  Contact us for more information or to request a demo today!

 


Discover more from Lucidity

Subscribe now to keep reading and get access to the full archive.

Continue reading